UCC MMI-2 troubleshooting

UCC MMI-2 visual diagnostics

UCC MMI-2 front panel

A visual indication of the system status is provided by two LEDs on the front panel, providing assistance in diagnosing and rectifying system faults.

LED status key


Description

SPA3-2 LED visual diagnostics - green yellow red

LED on, displaying green, amber or red

SPA3-2 visual diagnostics - Red flash

Red flash

SPA3-2 visual diagnostics - Green flash

Green flash

SPA3-2 visual diagnostics - no LED

LED off


Probe seated

Error

Description

Fault

Solution

SPA3-2 visual diagnostics - no LED

SPA3-2 LED visual diagnostics - red

Constant red error LED

Power fault/ FPGA reset

In the event of this fault, check the power supply or else, please return to Renishaw

SPA3-2 visual diagnostics - no LED


SPA3-2 visual diagnostics - Red flash

Intermittent fast flashing red error LED

Security validation failure

In the event of this fault, please return to Renishaw

SPA3-2 visual diagnostics - no LED


SPA3-2 visual diagnostics - Red flash

Continuous fast flashing red error LED

Factory configuration

Application FPGA image must be re-programmed and if fails, return to Renishaw

SPA3-2 visual diagnostics - Green flash

SPA3-2 visual diagnostics - Green flash

Slow green flashing error and probe LEDs

Normal start-up operation, waiting for comms synchronisation

Situation occurs during startup, software comms synchronisation or lost comms with software

SPA3-2 LED visual diagnostics - green

SPA3-2 visual diagnostics - no LED

Probe LED constant green

No fault

Normal operation

Probe seated, normal operation

SPA3-2 LED visual diagnostics - red

SPA3-2 visual diagnostics - no LED

Probe LED constant red

No fault

Normal operation

Probe open, normal operation


Fatal faults

Situations can occur that make it inadvisable or dangerous to continue using the CMM system. These are known in UCCsuite as fatal faults.

A list of UCC MMI-2 related fatal faults is shown below and will be indicated through the user's software (for example MODUS):

  • Air pressure is too low
  • A scale reading failure
  • An indicated overspeed (calculated from the rate of change of position)

NOTE: Other faults not classed as fatal can prevent the CMM's operation.

If the fatal fault cannot be resolved, please contact your local CMM support centre who can advise.